Instructions for Calibrating the iSavi
Terminal
© Addvalue Innovation Pte Ltd
8
13. Once the verification is completed, check that the message Calibration
completed. Result: Successful is displayed.
14. If the calibration process fails, all the four Antenna Pointing LEDs will
flash red. Check that the message
Calibration completed. Calibration failed! Please change your
existing location and recalibrate again is displayed.
Please repeat the procedure by clicking on the Recalibrate button.
PLEASE NOTE: Failure of the calibration can be caused by magnetometer
interference from the surrounding area. You need to change to a new location for
recalibration process.